Issue 5484: Control Flow (lecis-ftf) Source: Creon-Labcontrol AG (Mr. Thorsten Richter, thorsten.richter@creonlabcontrol.com) Nature: Uncategorized Issue Severity: Summary: The specification does not always define what happens in case I send a primary command to the SLM Main Unit (which would be delegated also to all Sub Units) but the main unit is not in the correct state. - The Main Unit should return the result code MAIN_STATE_INCORRECT and not call any Sub Units. T This is defined for some primary comamnds, but we should include a general statement into the spec to make this clear Resolution: Add explanatory paragraph on page 2-10 at the end of chapter 2.3.5 Revised Text: Text added: "In case a primary command is sent to the SLM Main Unit and the main unit is not in the correct state, the Main Unit returns the result code MAIN_STATE_INCORRECT and does not call any Sub Units." Actions taken: July 12, 2002: received issue December 11, 2002: closed issue Discussion: End of Annotations:===== To: lecis-ftf@omg.org, lecis-post@yahoogroups.com Subject: LECIS FTF issue TR13: Control Flow X-Mailer: Lotus Notes Release 5.0.5 September 22, 2000 From: Thorsten.Richter@creonlabcontrol.com Date: Fri, 12 Jul 2002 12:00:19 +0200 X-MIMETrack: Serialize by Router on domino1/creonlabcontrol/de(Release 5.0.8 |June 18, 2001) at 07/12/2002 12:02:27 PM, Serialize complete at 07/12/2002 12:02:27 PM The specification does not always define what happens in case I send a primary command to the SLM Main Unit (which would be delegated also to all Sub Units) but the main unit is not in the correct state. - The Main Unit should return the result code MAIN_STATE_INCORRECT and not call any Sub Units. T This is defined for some primary comamnds, but we should include a general statement into the spec to make this clear. Best regards, Thorsten Thorsten Richter Project Manager -------------------------------------------------------------- CREONLLLLLLLABCCCCCCCONTROL AG Europaallee 27-29, D-50226 Frechen, Germany Tel. +49 2234 9207-0 Fax. +49 2234 9207-99 http://www.creonlabcontrol.com -------------------------------------------------------------- X-Sender: tstaab@popmail.esa.lanl.gov X-Mailer: QUALCOMM Windows Eudora Version 5.1 Date: Tue, 06 Aug 2002 11:09:43 -0600 To: lecis-ftf@omg.org From: Torsten Staab Subject: LECIS FTF issue # TR13 In response to issue # TR13 Thorsten.Richter@creonlabcontrol.com wrote: Control Flow The specification does not always define what happens in case I send a primary command to the SLM Main Unit (which would be delegated also to all Sub Units) but the main unit is not in the correct state. - The Main Unit should return the result code MAIN_STATE_INCORRECT and not call any Sub Units. This is defined for some primary comamnds, but we should include a general statement into the spec to make this clear. Not forwarding invalid commands to the subunits makes a lot of sense. It simplifies the error handling, state recovery, and shortens the device's error response time. I'm in favor of adding a general statement to the spec that requires command filtering at the SLM main unit level. Torsten